Remote Operation

Introduction

3

3-3

Introduction

Two operating modes—local or remote—control the Analyzer. When the Analyzer is

powered on, it initializes in local mode. The Analyzer keyboard is enabled, and the

Analyzer can understand information input from its front panel.
The remote mode allows communication with the Analyzer, using a serial controller (a

computer with an RS232 port or a medTester with medCheck). Connect the Analyzer to

the controller with a null modem cable. As soon as the Analyzer receives a remote

command from a controller, it goes into the remote mode and recognizes commands from

only that controller.
The Analyzer keyboard is completely disabled during remote operation, except during

some procedures in which the

Esc

or function keys are needed. To return the Analyzer to

the local mode, use the

GOTOLOCAL

command. Program examples are given at the

end of this chapter.
There are two types of commands associated with the Analyzer:

Those that instruct the Analyzer to simply perform a procedure

Those that cause the Analyzer to perform a procedure and send test results (data)

back to the controlling device.

If the command instructs the Analyzer to do a procedure that generates no data, such as

drawing a pressure envelope, the Analyzer completes the procedure, then sends an

asterisk (*) to the controller. When the asterisk displays, the Analyzer has received,

understood, and carried out a command.
The following sections on remote commands include instructions on writing commands,

an error message list, a section of detailed command descriptions, and examples of how

to use the Analyzer commands in a computer program.

Setting Up the medTester

Before the Analyzer can communicate with a medTester, the medTester must be

configured, using the values listed in Table 3-1.

Table 3-1. medTester Remote Configuration Settings

medTester RS232

Setting

Com1 Off

Com2 On

Connect null modem cable from the Analyzer to
Com2.

Baud Rate

9600

Stop bits

1

Parity Off

CTS Off

.