Utilities – Fluke Biomedical Cufflink User Manual

Page 62

Advertising
background image

Cufflink

Operators Manual

2-34

Note

If an auto sequence is configured to print and is run before a printer is

properly connected to the Analyzer, a Printer Not Ready message

displays.If a printout is not desired, select

Edit

and make sure the

Print

BP Results

line says

NO

.

Utilities

The auto sequence utilities, accessible from the

Utility

submenu, lets the user edit, view,

name, and print auto sequences. To access any of the utilities described below from the

Utility

submenu (the

Auto Sequence Utilities

screen), highlight the desired utility

option and press the

Ent

key. The appropriate

screen displays.

Edit

The

Edit

utility lets the user change the settings of an auto sequence.

To modify the content of an auto sequence:
1. From the

Auto Sequence Utilities

screen, highlight

Edit

and select an auto

sequence with the function keys. The three Analyzer auto sequence editing pages are

shown below:

F1

F2

F3

F4

F5

Auto-1

Pressure Test. :

YES

Leak Test..... : YES
Pop Off Test.. : YES
Print BP Results YES

Store

No/Dec

PgDn

PgUp

Yes/Inc

fcv112.eps

F1

F2

F3

F4

F5

Auto-1 2

Blood Pressure HR (BPM) CYCLES

(1)

60/ 30 ( 40) Adult

40 1

(2) 80/ 50 ( 62) Adult

60 1

(3) 100/ 55 ( 75) Adult

80 1

(4) 120/ 80 ( 90) Adult

80 1

Store

No/Dec

PgDn

PgUp

Yes/Inc