Cuff deflation – Fluke Biomedical Cufflink User Manual

Page 141

Advertising
background image

Appendices

Non-Invasive Blood Pressure (NIBP) Monitoring Tutorial

A

A-5

Cuff Deflation

The linear deflate method is also known as continuous bleed. This is the method a nurse

typically uses when measuring blood pressure manually. If the deflate is slow, accuracy

is improved.
For example, if the bleed rate is 1 mmHg/sec and the heart rate is 60 BPM, the cuff

pressure changes 1 mmHg per heart beat. Therefore, the error due to change in cuff

pressure is limited to 1 mmHg. If the bleed rate is 10 mmHg/sec, then the cuff pressure

changes 10 mmHg per heartbeat, and the potential error due to change in cuff pressure is

10 mmHg. This error is in addition to the error of the method being used to determine

blood pressure. The tradeoff is accuracy versus patient discomfort.

Note

If someone taking a subject’s blood pressure deflates the cuff faster than

3 mmHg/sec, an appreciable error in the ability to measure blood pressure

may be introduced. For example, if the cuff is inflated to 180 mmHg and

then deflated to 60 mmHg, it should take 40 seconds at a deflation rate of

3 mmHg/second.

Automated monitors employ algorithms to interpret the measurement points to minimize

error due to rapid deflate. A typical deflate method used by automated monitors is known

as a step or stepwise deflate, as shown in Figure A-3.

150

100

Amplified P

u

lses

mmHg

mmHg

Step Deflate

P

u

lse Pair

50

2.0

1.0

Time

C

u

ff Press

u

re

P

u

lse Aplit

u

de

Rapid Deflate

fcv008.eps

Figure A-3. Cuff Pressure and Pulse Amplitude vs. Time: Commercial Monitor

The cuff is inflated to a pressure much greater than systole and deflated in discrete steps.

Typical step deflate size is from 4 to 10 mmHg depending on the monitor and cuff size

being used. After the deflate, the monitor measures some number of pulses before

Advertising