Cufflink

Operators Manual

3-10

Max

BP =

8

0/50, HR = 60 BPM

1000

Min

mm

7

1.0

0.

8

0.6

0.4

0.2

0

40

8

0

120

160

200

fcv038.eps

Returned

Data

*

Example

Draw a pressure envelope with a heart rate of 30 BPM, a blood pressure of 200/150, a 50% gain, and a
shift of 20 mmHg.

Command

drawenv

hr---30,bp=adams200/150,envgain=50,envshift=20

Results

The specified pressure envelope is drawn on the Analyzer display:

Max

BP = 200/150, HR = 30 BPM, gain = 50

%

, shift = 20 mmHg

500

Min

mm

63

1.0

0.

8

0.6

0.4

0.2

0

120

160

200

240

2

8

0

fcv039.eps

Returned

Data

*