CALPUCKPOS

Use this command to calibrate puck position. This procedure positions the puck at a

known location or reference point from which subsequent puck movement is calculated.

During this procedure, the Analyzer moves the puck in a downward direction until it

extends to its full range of movement (to the chamber bottom). Then the puck moves

up and stops as soon as it is detected by an opto sensor. The Analyzer records the

number of motor steps necessary to move the puck this distance and stores the number in

EEROM. There are no parameters for this command.

Example

Calibrate the puck position.

Command calpuckpos

Results

The Analyzer calibrates the puck. Returned data is

the value that

the Analyzer stores in EEROM

Returned Data

5681; varies with each Analyzer

DEFLATE

Use this command to release all pressure inside the Analyzer. There are no parameters for

this command.

Example

Eliminate the pressure inside the Analyzer.

Command deflate

Results

The Analyzer releases pressure.

Returned

Data

*

DRAWENV

Use this command to draw an ADAMS Adult pressure envelope on the Analyzer display,

as shown in Figure 3-1. Envelope gain, shift, and pulse width can be modified.

Max

Press

u

re En

v

elope

1000

Min

mm

1

1.0

0.

8

0.6

0.4

0.2

0

40

8

0

120

160

200

fcv037.eps

Figure 3-1. Analyzer Adult Blood Pressure Envelope (BP = 120/80)