Fluke Biomedical Cufflink User Manual
Page 87
Advertising

Remote Operation
Command Descriptions
3
3-13
Example
Draw a neonatal pressure envelope with a BP of 80/50 with a heart rate of 60 BPM.
Command
drawenvneo
hr=60,bp=adamsneo-80/50
Results
The specified pressure envelope is drawn on the
Analyzer display:
Max
BP =
8
0/50, HR = 60 BPM
1000
Min
mm
7
1.0
0.
8
0.6
0.4
0.2
0
40
8
0
120
160
200
fcv041.eps
Returned
Data
*
Example
Draw a neonatal pressure envelope with a heart rate of 30 BPM, a blood pressure of 150/100, a 50% gain,
and a shift of 20 mmHg.
Command
drawenvneo
hr=30,bp=adamsneo-150/100,envgain=50,envshift=-20
Results
The specified pressure envelope is drawn on the Analyzer display: