Remote Operation

Command Descriptions

3

3-13

Example

Draw a neonatal pressure envelope with a BP of 80/50 with a heart rate of 60 BPM.

Command

drawenvneo

hr=60,bp=adamsneo-80/50

Results

The specified pressure envelope is drawn on the
Analyzer display:

Max

BP =

8

0/50, HR = 60 BPM

1000

Min

mm

7

1.0

0.

8

0.6

0.4

0.2

0

40

8

0

120

160

200

fcv041.eps

Returned

Data

*

Example

Draw a neonatal pressure envelope with a heart rate of 30 BPM, a blood pressure of 150/100, a 50% gain,
and a shift of 20 mmHg.

Command

drawenvneo

hr=30,bp=adamsneo-150/100,envgain=50,envshift=-20

Results

The specified pressure envelope is drawn on the Analyzer display: