Cufflink

Operators Manual

2-12

F1

F2

F3

F4

F5

*** Pressure Curve Adjust ***

120/ 80 (90)

Draw

Shift

Gain%

fcv111.eps

1. Press

F1 Gain%

to move the selection box to

Gain%

and use the

Up

and

Down

arrow keys

to change the percent of gain. This factor affects the pressure pulse

amplitude. The default value is 100 %, and the range is from 1 % to 200 %. A change

in the Gain% appears below the heart rate on the Makearm graph:

C

u

ffPres

C

u

ffPeak

DeflRate
DeflTime
InflRate
InflTime
TotlTime
120/

8

0

0
162
5.1
17.5
46.7
3.3
22.0
(90) A

200

160

120

40

0

10

20

30

8

0 BPM

G 50
S 12

En

v

elope Gain

%

8

0

fcv020.eps

2. Press

F2 Shift

to move the selection box to

Shift

and use the

Up

and

Down arrow

keys

to shift the entire blood pressure envelope to the left (- shift,

Down arrow

) or

right (+ shift,

Up arrow

). If a target value of 120/80 is selected with a shift of +10

mmHg, the actual blood pressure target value simulated changes to 130/90. The

default value for a shift is 0 mmHg, and the range is from -100 mmHg to +100

mmHg.
The graphs below illustrate a blood pressure envelope with no shift, a negative shift,

and a positive shift, respectively: