The CuffLink Non-Invasive Blood Pressure Analyzer, hereafter referred to as the

Analyzer, provides accurate and repeatable dynamic blood pressure (BP) waveforms for

evaluation of both semi and fully automated oscillometric non-invasive blood pressure

(NIBP) devices. The Analyzer is shown in Figure 1-1.

F1

ESC

ENT

F2

F3

F4

F5

CuffLink

NON-INV

ASIVE BLOOD PRESSURE A

NALYZER

CUFF

CO

NN

ECT

CUFF

OUTPUTS

PULSE

BEEPER

V

OLUM

E

DISPL

AY

V

IE

W

Max

DC

mHg

mmHg

mmHg

Da

rk

fcv001.eps

Figure 1-1. CuffLink Non-Invasive Blood Pressure Analyzer

To test a device, wrap the BP cuff around the supplied mandrel and insert the cuff adapter

in the pressurized line. All tests are conducted with the BP cuff connected to the system.
The Analyzer can generate BP waveforms for seven adult (oscillometric), five neonate

(oscillometric), and 5 arrhythmias. The different systolic/diastolic pressure gradients

simulate a physiological range of normal, hypotensive, and hypertensive adult or neonate

patients. Actual patient data was used to design the preprogrammed peripheral pulse

waveforms and envelopes.
The Analyzer can produce qualitative measurements of BP cuff pressure and inflate /

deflate timing, as shown in Figure 1-2.