Powering up the analyzer, Menu structure and navigation – Fluke Biomedical Cufflink User Manual

Page 31

Advertising
background image

Operation

Powering Up the Analyzer

2

2-3

Powering Up the Analyzer

To power up the Analyzer, follow these directions:
1. Attach the supplied power cord to the power cord input on the back panel.
2. Plug the unit into a properly rated outlet.
3. Turn the Analyzer on by pushing the power switch on the back panel to the

ON

position (marked I). The Analyzer performs a self test and system initialization,

during which the logo display is visible for about five seconds:

F1

F2

F3

F4

F5

16:15:36
09/19/07
CuffLink

3.21, Pump

Non-Invasive Blood Pressure Analyzer

fcv101.eps

The current time and date is in the upper right hand corner, and the software revision,

along with installed options, appears directly below the Analyzer name. After the

five-second logo display, the

Main Menu

displays:

F1

F2

F3

F4

F5

Select BP

Press Util Comm Auto

ADAMS Adult

ADAMS Adult
ADAMS Neonate
Arrythmias

Select adult ADAMS family blood pressure

fcv096.eps

Menu Structure and Navigation

The

Main Menu

options listed across the top of the display are:

Select BP

(Select Blood Pressure)

Press

(Pressure Tests)

Util

(Utilities)

Comm

(Communications Ports)

Auto

(Auto Sequences)

Each of these options has its own submenu. See Figure 2-1 for a complete menu map.
Along the bottom of the

Main Menu

display is a brief description of the contents or

purpose of the highlighted menu option. If a pull down menu, or submenu, is visible

below the highlighted

Main Menu

option, the menu description refers to what is

available in the highlighted submenu.

Advertising