Pumppcb, Rcuserenv, Rdclock – Fluke Biomedical Cufflink User Manual

Page 110: Rdenvgain

Advertising
background image

Cufflink

Operators Manual

3-36

PUMPPCB

Use this command to determine if a pump circuit board is installed in the Analyzer. There

are no parameters for this command.

Example

Check for pump PCB.

Command pumppcb

Results

The analyzer scans for the board.

Returned

Data

Pump PCB installed or Pump PCB not installed

RCUSERENV

Use this command to copy the user envelope data from EEROM to RAM. There are no

parameters for this command.

Example

Recall user envelope data.

Command

rcuserenv

Results

Cufflink writes the user envelope data stored in EEROM to RAM.

Returned

Data

*

RDCLOCK

Use this command to display the current time, day, and date. The data returned to the

controller is configured as hour:min:sec mo/date/yr day (0 = Sunday, 6 = Saturday).

There are no parameters for this command.

Example

Display the time and date for Monday, December 5th.

Command

rdclock

Results

The Analyzer reads the clock information.

Returned Data

13:14:12 12/05/91 1; varies, depending on the current date

RDENVGAIN

Use this command to show the pressure envelope gain value (displayed in %) on the

controller's display. There are no parameters for this command.

Example

Display the envelope gain value.

Command rdenvgain

Results

reads the envelope gain value from ROM.

Returned Data

101; varies, depending on the current setting for envelope gain