Simulating adult blood pressure – Fluke Biomedical Cufflink User Manual

Page 42

Advertising
background image

Cufflink

Operators Manual

2-14

C

u

ffPres

C

u

ffPeak

DeflRate
DeflTime
InflRate
InflTime
TotlTime
120/

8

0

0
162
5.1
17.5
46.7
3.3
22.0
(90) A

200

160

120

40

0

10

20

30

8

0 BPM

G 50
S 12

En

v

elope Shift

8

0

fcv023.eps

3. Press the

Esc

key to undo any changes.

4. When finished, press the

Ent

key to save the pressure envelope and return to the

Pressure Curve Adjust

screen.

5. When finished, press

F3 Draw

to view the pressure envelope.

6. Press the

Ent

key.

The user is now ready to start the NIBP monitor, simulate the blood pressure, and take

measurements.

Simulating Adult Blood Pressure

To simulate the ADAMS Adult 120/80 Blood Pressure target value:
1. From the

Select BP

submenu, highlight

ADAMS Adult

and press the

Ent

key.

Target values and available blood pressures are displayed for the

ADAMS Adult

Family Target Values

screen. The cursor is initially at 120/80. The heart rate is

fixed at 80 BPM.

2. Press the

Ent

key to enter the Makearm function for a blood pressure of 120/80. The

Analyzer briefly displays a confirmation, verifying which parameter was chosen:

120/80 (90)

Note

The analyzer pop-up windows are not visible if Pop Time is set to 0.00

Advertising