Remote Operation

Programming with Analyzer Commands

3

3-41

ZEROPRESS

Use this command to zero the pressure transducer. This removes any offset errors during

a pressure measurement. To get an idea of how often pressure should be zeroed,

periodically make a pressure measurement with nothing connected to the Cuff Connect

output to see if the reading is zero. If it is not zero, issue to the Analyzer a zeropress

command. There are no parameters for this command.

Example

Zero the pressure transducer.

Command zeropress

Results

The Analyzer removes any offset occurring at the pressure transducer
and sets it to zero.

Returned

Data

*

Programming with Analyzer Commands

With the Analyzer remote commands, custom programs can be created that automate

calibration and testing protocols. This section contains samples of BASIC code,

checklists, and test results.

Checklist Generated in BASIC

The BASIC program shown in Figure 3-6 tells the Analyzer to display its firmware

version, simulate an adult blood pressure, display the test results on the computer's

display, and return to local mode.