Table 2-12. RS232 Settings

Parameter Value

Baud Rate

300

600

1200

2400

4800

9600

Bits/Character

7

8

Parity Mode

Even

Odd

None

Stop Bits

1

2

Flow Control

Xon/Xoff

RTS/CTS

None

3. If saving the new values is not desired, press the

Esc

key. If saving the new values is

desired, press the

Ent

key. In either case, the Analyzer returns to the

Comm

submenu.

Testing the RS232 Port and Connections

To test the RS232 port and its connections:
1. From the

Comm

submenu, highlight

Comm Test

and press the

Ent

key. The

Comm Test

screen displays:

F1

F2

F3

F4

F5

*** Comm port test ***
Baud, Len, Parity,Stop: 9600,8,none,1
Flow Control: XON/XOFF
Transmit Receive
Hit F2 to send Hit F4 to clear
CuffLink Test this box
Transmit Clear

fcv108.eps

At the top of the screen are the current settings for the RS232 port.

2. Press

F2 Transmit

to transmit a signal to whatever is being used as the controller.

The phrase Analyzer Test should appear on the controller's monitor.