Remote Operation

Command Descriptions

3

3-7

Command Descriptions

This alphabetized section contains detailed descriptions of the Analyzer's remote

commands. Examples for each command show exactly how to type the command, the

results, and the data returned to the controller, as it appears on the controller's display.
Commands in the Examples sections were written in command syntax for computers. For

information on medTester syntax, refer to Command Syntax for medTester.

Note

Different results from those shown in the examples may be obtained. Except for asterisks,

the Returned Data listed in the following examples is varied, depending on the Analyzer

and the parameter values assigned to the command.

BEEP

Use this command to test Cufflink's speaker and modify the frequency (pitch) and period

(length of the beep). Because the speaker is more difficult to hear at very low and very

high frequencies, be sure the Beeper Volume control on the front panel is set high

enough. Parameters for this command are described below.

Parameter Definition

Range

Default

FREQ

Frequency (Hertz)

76 to 10,000 Hz

440 Hz

PERIOD

Period (milliseconds)

50 to 32,000 ms

500 ms

Example

Beep the speaker using Cufflink's default values.

Command

beep

Results

A 440 Hz tone sounds for 500 milliseconds.

Returned

Data

*

Example

Beep the speaker at a frequency of 300 Hz for 800 milliseconds.

Command

beep

freq=300,period=800

Results

A 440 Hz tone sounds for 500 milliseconds.

Returned

Data

*