Remote Operation

Command Descriptions

3

3-19

Example

Pump Analyzer pressure up to 200 mmHg.

Command inflate

Results

The Analyzer internal pump increases pressure to 200 mmHg.

Returned

Data

*

KEYTEST

Use this command to test the Analyzer keyboard. There are no parameters for this

command.

Example

Test the keyboard.

Command keytest

Results

The Analyzer echoes the key pressed. To exit the keytest, press Esc
four times.

Returned

Data

F1

LEAKTEST

Use this command to put the Analyzer in the leak test mode. The keys on the Analyzer

front panel become active during this test so a user can zero pressure, perform the leak

test, do a reset, or print the test results. To exit the leak test mode, simply issue to the

Analyzer another remote command. There are no parameters for this command.

Example

Enter the Analyzer leak test mode.

Command leaktest

Results

The Analyzer enters the leak test mode and shows the Leak Test
display. Press F1 on the keyboard to begin the leak test. The leak test
runs for one minute and then displays the leak rate in mmHg/min.

F1

F2