Fluke Biomedical Cufflink User Manual

Page 26

Advertising
background image

Cufflink

Operators Manual

1-14

Preprogrammed Pulse Envelope

Horizontal Axis (cuff pressure) ............................... 0 to 300 mmHg in 1.0 mmHg steps
Vertical Axis ............................................................ 2.0 mmHg (nominal)

100% gain with normal adult cuff
0 to 2% selectable

Pulse Amplitude (adult) .......................................... 2.0 mmHg @ MAP 100% gain
Repeatability ...........................................................

±

1 % of selected target value

Pulse Waveforms

Pulse ID #

Pulse Width (ms)

Rise Time (ms)

0 800

270

1 500

165

2 250 85

3 720 90

4 230 80

5 280 96

6 350

100

7 480

108

8 980

180

9 1980

460

10 1480 330

Digital Manometer

Pressure Range ...................................................... Maximum = 499.75 mmHg
Measurement Parameters ...................................... Instantaneous and peak
Pump....................................................................... 2.0 liter/minute minimum (free flow)

Automated Leak Testing

Start Pressure ......................................................... Maximum = 499.75 mmHg
Elapsed Time .......................................................... Fixed at 60 seconds
Leak Rate Range .................................................... 0.25 to 499.75 mmHg/minute
Pump....................................................................... 2.0 liter/minute minimum (free flow)

Monitor Pop-off Relief Valve Testing

Measurement Parameters ...................................... Instantaneous and peak pressure
Maximum Pressure ................................................. 499.75 mmHg

Accuracy

Dynamic NIBP Response Repeatability
(Systolic/Diastalic/Mean).........................................

±

1 % of Target Value

Cuff Pressure ..........................................................

±

1.0 % of reading (±1 mmHg)

Input Overpressure Limit.........................................

±

1500 mmHg