Mkarr_pac, Mkarr_pvc, Mkarr_as – Fluke Biomedical Cufflink User Manual

Page 105

Advertising
background image

Remote Operation

Command Descriptions

3

3-31

MKARR_PAC

Use this command to set blood pressure simulation to the Premature Atrial Contraction

Arrhythmia at the fixed BP and heart rate of 120/80 and 80 BPM. There are no

parameters for this command.

Example

Simulate a premature atrial contraction.

Command mkarr_pac

Results

The Analyzer runs premature atrial contraction at 120/80 with the
heart rate set to 80 BPM

Returned

Data

*

MKARR_PVC

Use this command to set blood pressure simulation to the Premature Ventricular

Contraction Arrhythmia at the fixed BP and heart rate of 120/80 and 80 BPM. There are

no parameters for this command.

Example

Simulate premature ventricular contraction.

Command mkarr_pvc

Results

The Analyzer runs premature ventricular contraction at 120/80 with
the heart rate set to 80 BPM.

Returned

Data

*

.

MKARR_AS

Use this command to set blood pressure simulation to the Aberrant Sinus Conduction

Arrhythmia at the fixed BP and heart rate of 120/80 and 80 BPM. There are no

parameters for this command.

Example

Simulate aberrant sinus conduction.

Command mkarr_as

Results

The Analyzer runs premature aberrant sinus conduction at 120/80
with the heart rate set to 80 BPM.

Returned

Data

*