Fluke Biomedical Cufflink User Manual

Page 55

Advertising
background image

Operation

Utility Functions

2

2-27

Table 2-8. Speaker Tests and Adjustments

Parameter Procedure

440 Press

F1; the speaker produces a 440 hertz, 500 millisecond tone

Adjust Play

Press F2, the speaker produces a tone determined by the settings determined by
the Adjust Frequency and Adjust Period values (see below for description)
shown in the center of the display.

Adjust Freq

Press F3; the cursor moves to the Adjust Frequency value. When the cursor is at
this position, the frequency of the Adjust Play tone may be modified. Use the Up
or Down arrow keys to change the value. The Left and Right arrow keys have no
effect on this function. Change the value of the frequency and then press F2
(Adjust Play) to hear the difference in the pitch of the tone. As the value of the
frequency increases, so does the pitch.

Adjust
Period

Press F4; the cursor moves to the Adjust Period value, displayed in milliseconds
(ms). When the cursor is at this position, the length of the Adjust Play tone may be
modified. Use the Up or Down arrow keys to change the value. The Left and
Right arrow keys have no effect on this function. Change the value of the period
and press F2 (Adjust Play) to hear the difference in length of the tone. As the
value of the period increases, so does the length

Sweep Press

F5; the speaker produces a long tone of various frequencies, starting with a

low frequency (pitch), gradually increasing to a higher frequency, then returning to
the low frequency from which it began.

3. To exit the

Speaker Test

option, press the

Esc

key.

Display Test

The Display Test verifies the operation of the Analyzer liquid crystal display by writing

(displaying) a series of test patterns to the display. Any sections of the display that are not

functioning correctly become apparent during this test.
To carry out the Display Test:
1. On the

System Utilities

screen, highlight the

Display Test

option and press the

Ent

key. The

Display Test

screen displays.

2. Test the various parameters as desired, according to the instructions in Table 2-9.