Gotolocal, Ident, Inflate – Fluke Biomedical Cufflink User Manual

Page 92

Advertising
background image

Cufflink

Operators Manual

3-18

Example

Do a global initialization.

Command globalinit

Results

All user modifications for Pop Time, RS232 port, and autosequences
are reset to factory defaults.

Returned

Data

*

GOTOLOCAL

Use this command to switch control of the Analyzer fro a serial controller to the Analyzer

keyboard. There are no parameters for this command.

Example

Return the Analyzer control to its front panel.

Command gotolocal

Results

The Analyzer logo displays, and it returns to the local mode. The front
panel keyboard is now active.

Returned

Data

*

IDENT

Use this command to show the Analyzer name and firmware version on the controller

display. There are no parameters for this command.

Example

Identify the Analyzer and firmware version.

Command ident

Results

Display of Analyzer identification and firmware version

Returned Data

CuffLink, 3.00, Pump; varies with each Analyzer, depending on the
installed firmware and options

INFLATE

Use this command to pump up the pressure inside the Analyzer. Parameters for this

command are described below.

Parameter Definition

Range

Default

PRESSURE

Pressure (mmHg)

0 to 499.75 mmHg

200 mmHg

Example

Pump Analyzer pressure up to 275 mmHg.

Command Inflate

pressure=275

Results

The Analyzer internal pump increases pressure to 275 mmHg.

Returned

Data

*