Makearm – Fluke Biomedical Cufflink User Manual

Page 94

Advertising
background image

Cufflink

Operators Manual

3-20

MAKEARM

Use this command to simulate blood pressures from the ADAMS Adult Family target

values. Heart rate, BP, and pressure envelope gain and shift can be modified. Parameters

for this command are described below.

Parameter Definition

Range Default

Description

HR Heart

Rate

(Beats/Minute)

30 BPM to 240
BPM in 1 BPM
Steps

80 BPM

1 BPM resolution

BP Blood

Pressure

(mmHg)

ADAMS-60/30

ADAMS-80/50

ADAMS-100/65

ADAMS-120/80

ADAMS-
150/100

ADAMS-
200/150

ADAMS-
255/195

ADAMS-120/80

PULSEID Pulse

Identification
(milliseconds)

ID# / WIDTH

0 / 800 ms

1 / 500 ms

2 / 250 ms

3 / 720 ms

4 / 230 ms

5 / 280 ms

6 / 350 ms

7 / 480 ms

8 / 980 ms

9 / 1980 ms

10 / 1480 ms

ID# 3
(720 ms)

Pulse width affects
the rise time of
pulse

ENVGAIN Pressure

Envelope Gain

1 to 200%
(10 to 2000
RVDUs)

100% is a typical
value for adult
cuffs.

Affects pulse
strength or
amplitude

ENVSHIFT Pressure

Envelope Shift

-100 to +100
mmHg

0 mmHg

Modifies existing
pressure envelope
to simulate
another BP

Example: 0 +10
mmHg shift
causes 120/80 to
read 130/90 on
most monitors.