Readpress, Reset – Fluke Biomedical Cufflink User Manual

Page 112

Advertising
background image

Cufflink

Operators Manual

3-38

Example

Read the user envelope

Command rduserenv

Results

The Analyzer takes data for the user envelope in RAM and sends it
out the RS232 port.

Returned Data

1:100. 2:100, 3:100, 4:105, 5:110 …. 298:150, 299:140, 300:130, *

READPRESS

Use this command to show a value (in 0.25 mmHg increments) on the controller's display

for the Analyzer pressure transducer. There are no parameters for this command.

Example

Read the pressure transducer value.

Command readpress

Results

The Analyzer reads the pressure transducer value stored in EEROM.

Returned Data

123.25; varies for each Analyzer

RESET

Use this command to reset the Analyzer back to factory default values. This affects the

parameters listed below. There are no parameters for this command.

Parameter Default

Value

Pulse ID number

3

Heart Rate

80 BPM

Pressure Curve ID number

A120 80

Pressure Curve Vertical Gain

100%

Pressure Curve Horizontal Shift

0 mmHg

Example

Reset CuffLink to default values

Command Reset

Results

The Analyzer resets to factory values.

Returned

Data

*