Pressure testing – Fluke Biomedical Cufflink User Manual

Page 17

Advertising
background image

Introduction and Specifications

New Features (Firmware Revision 3.0 and Later)

1

1-5

Premature Atrial Contraction (PAC)

The first pulse of the PAC cycle is premature and lower in amplitude than a normal sinus

pulse. The next pulse would be back in sync with normal sinus and slightly higher in

amplitude. All subsequent pulses are normal.

Premature Ventricular Contraction (PVC)

This is a representation of the peripheral pulse similar to PAC

Atrial Fibrillation (AF)

The AF cycle has an irregular R to R interval. Its occurrence and properties (early vs.

late) are random throughout the pressure curve cycle.

Missed Beat (MB)

A complete beat is randomly skipped during the pressure curve cycle. The following beat

reverts to normal R to R intervals.

Aberrant Sinus Conduction (AS)

The AS cycle inserts one pulse so low that it is virtually non-existent. This causes the

Analyzer to skip one diastolic pulse and then return to normal sinus pulses.

Pressure Testing

The Analyzer now has the following enhancements to its pressure testing capability.

Internal Pump

The Analyzer now has an internal compressor which eliminates the need to manually

inflate the cuff for NIBP monitor testing. This automates static pressure measurements,

leak testing, and relief valve testing.

Pop Off Added to Press Menu

Perform over-pressure tests on NIBP monitors with this addition to the

PRESS

menu.

Utility Menu Added to Leak Test

Select cuff size, turn printing on or off, and choose a target pressure for leak testing.

Manometer

The Analyzer simulates a digital manometer with pump capabilities.