Fluke Biomedical Cufflink User Manual

Page 128

Advertising
background image

Cufflink

Operators Manual

3-54

MedTester REC # 20

CHECKLIST: NIBP*ARR

OP CODE: TEO

DEVICE INFORMATION

PHYSICAL INSPECTION

COMMENTS:

NEXT TEST DUE DATE:

USER TIME:

ELAPSED TEST TIME: 208 SECS

GENERAL CONDITION OF EQUIPMENT........PASS

LINE CORD/STRAIN RELIEF...............PASS

TRPN (POWER ON).......................REMOTE CMD, NO DATA

INDICATORS/DISPLAYS...................PASS

CONTROLS/SWITCHES.....................PASS

TUBES/HOSES...........................PASS

FITTINGS/CONNECTORS...................PASS

IDENT.................................CUFFLINK,3.00,AUSC,PUMP

LEAKTEST..............................*

LEAK TEST: APPLY 200 MMHG (<10/MIN)...PASS

MANOMETER.............................*

STATIC PRESSURE CHECK: APPLY 200 MMHG.PASS

STATIC PRESSURE CHECK: APPLY 100 MMHG.PASS

POP OFF/RELIEF VALVE: APPLY >250 MMHG.PASS

MKARR_PAC PRE. ATRIAL CONTRACTION.....*

ARRHYTHMIA PAC 120/80 @ 80 BPM........PASS

MKARR_PVC PRE VENTRICULAR CONTRACTION*

ARRHYTHMIA PAC 120/80 @ 80 BPM........PASS

MKARR_AF ATRIAL FIBRILLATION..........*

ARRHYTHMIA ATRIAL FIB 120/80 @ 80 BPM.PASS

GOTOLOCAL.............................REMOTE CMD, NO DATA

GTOL..................................REMOTE CMD, NO DATA

TYPE: NIBP*ARR

MODEL:

MANF:

SN:

LOC: DNI

CN: NIBP*ARR

DATE: 6/05/93

TIME: 15:49:44

fcv077.eps

Figure 3-18. Sentinel Adult Arrhythmic BP Test Results