Pulse amplitude – Fluke Biomedical Cufflink User Manual

Page 140

Advertising
background image

Cufflink

Operators Manual

A-4

Pulse Amplitude

The amplitude of the oscillometric pulses (hereafter referred to as pulses) is quite small

when compared to the static pressure in the cuff. As shown in Figure A-2, these pulses

appear as very small spikes on the cuff pressure waveform. They are depicted in

amplified form with the cuff pressure stripped off to reveal how the amplitude varies as a

function of cuff pressure. The peak pulse amplitude is 2 mmHg at a cuff pressure of

115 mmHg in the example shown.

mmHg

P

u

lse Amplit

u

de

C

u

ff Press

u

re

mmHg

P

u

lse

En

v

elope

200

C

u

ff Press

u

re

Time

100

2.0

1.0

fcv007.eps

Figure A-2. Pulse Amplitude and Cuff Pressure vs. Time: Linear Deflate

In general, the peak pulse amplitude is 1-3 % of the cuff pressure at which it occurs.

Therefore, the monitor must be able to strip off the large static cuff pressure to measure

the individual pulse. Because the pulses are so small, it is possible for artifact conditions

to obscure the pulse. Patient motion and respiration are common artifacts that the monitor

must reject. When the cuff pressure is quite high the pulse amplitude is small. As the cuff

deflates, the pulse amplitudes increase to a maximum and then decrease to a minimum.

Advertising