Drawpulse – Fluke Biomedical Cufflink User Manual
Page 89
Advertising

Remote Operation
Command Descriptions
3
3-15
DRAWPULSE
Use this command to draw an individual pressure pulse, as shown in Figure . Different
pulse widths can be specified for this command.
msec
P
u
lse # 3 (720 milliseconds)
716
sec
1.0
0.
8
0.6
0.4
0.2
0
0.2
0.4
0.6
0.
8
1.0
fcv043.eps
Figure 3-3. Analyzer Pressure Pulse #3
Parameters for this command are described below.
Parameter Definition
Range Default
Description
PULSEID Pulse
Identification
(milliseconds)
ID# /
WIDTH
0 / 800 ms
1 / 500 ms
2 / 250 ms
3 / 720 ms
4 / 230 ms
5 / 280 ms
6 / 350 ms
7 / 480 ms
8 / 980 ms
9 / 1980 ms
10 / 1480
ms
ID# 3
(720
ms)
Pulse width affects the rise time
of pulse