Globalinit – Fluke Biomedical Cufflink User Manual

Page 91

Advertising
background image

Remote Operation

Command Descriptions

3

3-17

GLOBALINIT

Use this command to reset the Analyzer to factory defaults. There are no parameters for

this command. The default values are listed below.

Table 3-3. Autosequence Defaults

Cycles *

BP HR

A1 A2 A3 A4 A5

60/30

(40)

Adult

40 1 2 3 4 5

80/50

(62)

Adult

60 1 2 3 4 5

100/65

(75)

Adult

80 1 2 3 4 5

120/80

(90)

Adult

80 1 2 3 4 5

150/100

(115)

Adult

120

1 2 3 4 5

200/150

(165)

Adult

160

1 2 3 4 5

255/195

(215)

Adult

200

1 2 3 4 5

120/80

(80)

Adult

80 1 2 3 4 5

Utilities

Pressure

Test

Yes

Leak

Test

Yes

Pop

Off

Test

Yes

Print

BP

Results

Yes

* Cycles refer to the number of time that the Analyzer simulates a target value.

Table 3-4. Pop Time and Comm Port Default Values

Parameter Value

Pop Time

1 second

Communications

Baud Rate

9600

Bits/Character 8

Parity None

Stop Bits

1

Flow Control

Xon/Xoff