Operation

Printing Documents

2

2-37

Init

The

Init

utility lets the user reset an auto sequence to factory default values.

To reset an auto sequence to factory default values:
1. From the

Auto Sequence Utilities

screen, highlight

Init

and select an auto

sequence with the function keys. The Analyzer warns that the auto sequence is about

to be reset.

2. Press the

Ent

key to reset the values or press the

Esc

key to return to the

Utilities

submenu without resetting.

Print All

The

Print All

utility lets the user print the content of all auto sequences.

To print the content of all auto sequences:
1. From the

Auto Sequence Utilities

screen, highlight

Print All

and press the

Ent

key. The Analyzer displays Printing all Auto Sequences while the documents are

printing.

Init All

The

Init All

utility lets the user reset all auto sequences to factory defaults.

To reset all auto sequences to default values:
1. From the

Auto Sequence Utilities

screen, highlight

Init All

and press the

Ent

key. The Analyzer warns that all auto sequences are about to be reset.

2. Press the

Ent

key to reset all values to the default or press the

Esc

key to return to

the

Auto Sequence Utilities

screen without resetting. Refer to Figure 2-11 for a

list of auto sequence defaults.

Printing Documents

Test results can be printed from the Analyzer to any parallel printer suitable for use with

a PC-compatible computer. Connect the printer cable to the port labeled

Printer

on the

Analyzer rear panel.

Note

If the user tries to print a document when the printer is the wrong type, out

of paper, not on line, or not properly connected, the Analyzer displays an

error message stating Printer not ready.

Printing Blood Pressure Test Results

To print the results of blood pressure tests:
1. Use the arrow keys to choose one of the selections under

Select BP

on the Analyzer

Main Menu

and press the

Ent

key.

2. Press

F3 Print

to enable printing. A Printing enabled message displays.

3. Choose the blood pressure value to simulate and press the

Ent

key.

4. Start the NIBP monitor to begin the test. The Analyzer sends the test results to the

printer when the cuff has completely deflated. Figure 2-7 is a sample printout.