Pscale – Fluke Biomedical Cufflink User Manual

Page 107

Advertising
background image

Remote Operation

Command Descriptions

3

3-33

PSCALE

Use this command to produce a constant pulse amplitude that is independent of pressure,

via the Scale Pulse Amplitude display shown in Figure 3-5. This signal is present at all

Analyzer front panel outputs; the signal can be monitored at the PULSE output. The

Analyzer keyboard is enabled during this test. Heart rate, pulse width, pulse amplitude,

and puck motion can be modified.

F1

F2

F3

F4

F5

*** Scale Pulse Amplitude ***

Motion : Stopped Enter toggles
Amplitude : 1000 motion start/
Pressure

: 0.00 stop

Rate (BPM) 80

Rate

Shape 250

500

1000

fcv058.eps

Figure 3-5. Scale Pulse Amplitude Display

Parameters for this command are described below.

Parameter Definition

Range

Default

HR Heart

Rote

(Beats/Minute)

30 BPM to 240 BPM
in 1 BPM Steps

80 BPM
1 BPM resolution

PULSEAMP Pulse

Amplitude

(RVDU)

1 to 1000 RVDUs

1000 RVDUs

MOTION Puck

Motion

On
Off

On

PULSEID

Pulse Identification

ID# / WIDTH

0 / 800 ms
1 / 500 ms
2 / 250 ms
3 / 720 ms
4 / 230 ms
5 / 280 ms
6 / 350 ms
7 / 480 ms
8 / 980 ms
9 / 1980 ms
10 / 1480 ms

Pulse #3
(720 ms)

*RVDU = Relative Volume Displacement Unit, or the volume display when the pressure engine
moves one increment (one microstep).