Operation

Utility Functions

2

2-25

Logo

The

Logo

option shows the Analyzer name, current software revision, time, and date.

To view the logo:
1. From the

Util

submenu, highlight

Logo

and press the

Ent

key. The logo is

displayed:

F1

F2

F3

F4

F5

16:15:36
09/19/07
CuffLink

3.21, Pump

Non-Invasive Blood Pressure Analyzer

fcv101.eps

2. Press the

Esc

key to return the display to the

Util

submenu.

Note

No adjustments are possible in the Logo option.

System Functions

The

System

functions, accessed from the

Util

submenu, are primarily functional self

tests, available from the System Utilities screen:

F1

F2

F3

F4

F5

*** System Utilities ***

Print Test

ROM

Checksums

Key Test

Config Init

Speaker Test

User Envelope

Display Test

Pulse

Send test string to printer

fcv110.eps

Print Test

The Print Test verifies that the Analyzer can communicate with a connected printer.
To carry out the Print Test:
1. Connect the printer to the Analyzer printer port on the rear panel and power on the

printer.

2. On the

System Utilities

screen, highlight the

Print Test

option and press the

Ent

key. The Analyzer sends a test string through the printer port to the printer. If the

Analyzer is communicating correctly, the printer should print the following each time

the

Ent

key is pressed:

Printer test message