Fluke Biomedical Cufflink User Manual

Page 109

Advertising
background image

Remote Operation

Command Descriptions

3

3-35

Parameter Definition

Range

Default

PULSEAMP Pulse

Amplitude

(RVDU)

1 to 1000 RVDUs

1000 RVDUs

PULSEID

Pulse Identification

ID# / WIDTH

0 / 800 ms
1 / 500 ms
2 / 250 ms
3 / 720 ms
4 / 230 ms
5 / 280 ms
6 / 350 ms
7 / 480 ms
8 / 980 ms
9 / 1980 ms
10 / 1480 ms

Pulse #3
(720 ms)

Pulse width affects the
rise time of the pulse.

*RVDU = Relative Volume Displacement Unit, or the volume display when the pressure engine
moves one increment (one microstep).

Example

Output a pulse using the active pressure envelope and pulse width (ID).

Command pulse

Results

No parameters were specified in this example, so Cufflink determines
amplitude by reading cuff pressure and applies it to the active
pressure envelope. Cufflink also uses the current pulse width and
drives the motor to output a single pulse and then stop. Pulse is
output at PULSE output jack on the Cufflink front panel.

Returned

Data

*

Example

Output a pulse that is 231 steps in amplitude and 23 ms wide.

Command

pulse

pulseamp=231,pulseid=4

Results

Pulse #4 becomes the active pulse. If another pulse command is
issued without specifying parameters, the Analyzer uses pulseid #4.
Pulse is output at the PULSE output jack on the Analyzer front panel.

Returned

Data

*