Fluke Biomedical Cufflink User Manual

Page 123

Advertising
background image

Remote Operation

Programming with Analyzer Commands

3

3-49

DEMO ROOM

Page: 1

Date: 06/05/93

Time: 12:38 AM

Proc. Name

SENT*NEO

Description

GENERAL CONDITION OF EQUIPMENT

LINE CORD/STRAIN RELIEF

TRPN (POWER ON)

INDICATORS/DISPLAYS

CONTROLS/SWITCHES

TUBES/HOSES

FITTINGS/CONNECTORS

IDENT

LEAKTEST

LEAK TEST: APPLY 200 MMHG (<10/MIN)

MANOMETER

STATIC PRESSURE CHECK: APPLY 200 MMHG

STATIC PRESSURE CHECK: APPLY 100 MMHG

POP OFF/RELIEF VALVE: APPLY >250 MMHG

MAKEARMNEO:HR=60;BP=ADAMSNEO-150/100

ADAMS NEONATE 150/100 AT 60 BPM

MAKEARMNEO:HR=120;BP=ADAMSNEO-100/65

ADAMS NEONATE 100/65 AT 120 BPM

MAKEARMNEO:HR=200;BP=ADAMSNEO-60/30

ADAMS NEONATE 60/30 AT 200 BPM

GOTOLOCAL

GTOL

5

GEN/PHYS

GEN/MECH

REMOTE

GEN/ELEC

GEN/ELEC

GEN/PLMB

GEN/PHYS

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

AUTO SEQ

Pass/Fail

Pass/Fail

medTester remote command

Pass/Fail

Pass/Fail

Pass/Fail

Pass/Fail

medTester remote command

medTester remote command

Pass/Fail

medTester remote command

Pass/Fail

Pass/Fail

Pass/Fail

medTester remote command

Pass/Fail

medTester remote command

Pass/Fail

medTester remote command

Pass/Fail

medTester remote command

medTester remote command

medTester autosequence

Type

medTester Control

0.00

PM Labor Time

Relational Checklist Report

Listed In Order Of Relational Procedure

fcv073.eps

Figure 3-13. Sentinel Neonate BP Checklist

Advertising