Mkarm70, Mkarmneo10, Mkarmneo20 – Fluke Biomedical Cufflink User Manual

Page 103: Mkarmneo30

Advertising
background image

Remote Operation

Command Descriptions

3

3-29

MKARM70

Use this command to set blood pressure simulation to Adams Adult 255/195, with a heart

rate of 200 BPM. There are no parameters for this command.

Example

Simulate a specific adult BP of 255/195 at 200 BPM.

Command Mkarm70

Results

The Analyzer runs the Adams Adult 255/195, heart rate is set to 200
BPM.

Returned

Data

*

MKARMNEO10

Use this command to set blood pressure simulation to Adams Neonate 60/30, with a heart

rate of 40 BPM. There are no parameters for this command.

Example

Simulate a specific neonate BP of 60/30 at 40 BPM.

Command mkarmneo10

Results

The Analyzer runs the Adams Neonate 60/30, heart rate is set to 40
BPM.

Returned

Data

*

MKARMNEO20

Use this command to set blood pressure simulation to Adams Neonate 80/50, with a heart

rate of 60 BPM. There are no parameters for this command.

Example

Simulate a specific neonate BP of 80/50 at 60 BPM.

Command mkarmneo20

Results

The Analyzer runs the Adams Neonate 80/50, heart rate is set to 60
BPM.

Returned

Data

*

MKARMNEO30

Use this command to set blood pressure simulation to Adams Neonate 100/65, with a

heart rate of 80 BPM. There are no parameters for this command.

Example

Simulate a specific neonate BP of 100/65 at 80 BPM.

Command mkarmneo30

Results

The Analyzer runs the Adams Neonate 100/65, heart rate is set to 80
BPM.

Returned

Data

*