Cufflink

Operators Manual

3-40

Example

Send a clock information data string to the Analyzer.

Command wrclock

Results

A data string is sent to the Analyzer.

Returned

Data

*

WRPEAKDIV

Use this command to store a new peak divide ratio in EEROM. Peak divide ratio is the

variable used to detect the point of rapid cuff deflate or when to stop blood pressure

simulation during the makearm process. Set a divide ratio factor for this command.

Parameters for this command are described below.

Parameter Default

Value

Range

DIV

Divide Ratio Factor

2.0 to 20.0

Example

Set the peak divide ratio to 8.8.

Command wrpeakdiv

div=8.8

Results

The Analyzer stores a value of 8.8 in RAM for peak divide ratio.

Returned

Data

*

WRUSERENV

Use this command to write new user envelope data from Cufflink's RS232 port to RAM.

This action overwrites user envelope data already in RAM. Information can be written to

RAM (not to EEROM) only by using the wruserenv command. To write data to EEROM,

which is permanent, use the stuserenv command.
When the Analyzer receives a wruserenv command, it acknowledges that it is ready to

receive data by sending a star character (*) back to the controller. The controller can

begin sending the data, one line at a time, until all 300 lines have been received by the

Analyzer. The Analyzer marks the end of the transmission with another star character. A

PC utility program (PC3CUFF) is available that transfers the array text file from a 1PC to

the Analyzer. There are no parameters for this command.

Example

Write a new user envelope to RAM.

Command wruserenv

Results

The Analyzer writes new user envelope data from its RS232 port to
RAM.

Returned Data

1:100. 2:100, 3:100, 4:105, 5:110 …. 298:150, 299:140, 300:130, *