Standard features, New features (firmware revision 3.0 and later), Arrhythmias – Fluke Biomedical Cufflink User Manual

Page 16

Advertising
background image

Cufflink

Operators Manual

1-4

CUFF

PRESSURE

SYS

MAP

DIA

OSCILLATIO

N

AMPLITUDE

TIME

fcv002.eps

Figure 1-2. Cuff pressure waveform during blood pressure measurement

The Analyzer also offers automated leak testing of NIBP monitors. An internal pump

pressurizes the NIBP system under test. Press a key to initiate a 60 second leak test once

the desired pressure is reached. Use the Analyzer’s digital manometer instead of a

mercury column for doing pressure measurements. The Analyzer facilitates overpressure

testing of NIBP monitors by automatically detecting and displaying the overpressure

point.

Standard Features

The Analyzer has the following standard features:

Dynamic oscillometric noninvasive blood pressure simulation

Automated static pressure measurements, leakage testing, and relief-valve testing

Five automated NIBP testing autosequences

Five arrhythmia selections

Adult and neonatal NIBP selections

Direct interface with medTester 5000C

Adjustable heart rate values

Calendar clock with battery backup

Internal PCB expansion slot

New Features (Firmware Revision 3.0 and Later)

Firmware revision 3.0 included the following additions and updates, including a set of

adult arrhythmias, an internal pump, and additions to pressure testing.

Arrhythmias

The Analyzer features five new arrhythmias to test NIBP monitors in the presence of

typical patient arrhythmias. These clinically derived simulations are representations of

the peripheral pulse, as seen by an oscillometric NIBP monitor.
Each arrhythmia is generated on a random basis throughout the entire pressure curve

cycle. The variations in pulse timing and amplitude are relatively small.