Specifications – Fluke Biomedical Cufflink User Manual

Page 24

Advertising
background image

Cufflink

Operators Manual

1-12

Table 1-4. Rear Panel Controls and Indicators

Label Component

Description

H

Serial Port

The four digit Cufflink serial number is located above
the printer port. The serial number should be
documented along with the model number whenever
Cufflink is shipped to Dynatech Nevada.

I

Fuse Label

The fuse label documents the type of fuses needed.
Use one 1ASB 250 V ac fuse (DNI part no. 1005-
0184) if Cufflink is set for 120 V operation, and two
1/2 A 250 V ac fuses (DNI part no. 1005-0185) if
Cufflink is set for 240 V operation.

Replacing Fuses

1.

Turn Cufflink's power off and unplug the power cord.

2.

Remove the fuse cover with a small blade screwdriver.
The plastic fuse holder should pop out of the Cufflink
case.

3.

Use the screwdriver to pry the old fuse out of the plastic
holder.

4.

Install the new fuse.

5.

Replace the fuse holder by simply pushing the fuse
holder back into place.

Specifications

The following are general and electrical specifications for the Analyzer.

Physical Dimensions

Size ......................................................................... Width 12.5 inches

Height 5.0 inches
Length 15.0 inches

Weight..................................................................... 15 pounds

Power Requirements

Power ..................................................................... 120/250 V ac

50 Watts average
100 Watts peak
50/60 Hz

Input voltage range...................................................... 60 VA
Fuses ...................................................................... T3.15a 110-115 V ac 50/60 Hz

T1A 200-300 V ac 50/60 Hz

Environmental Conditions

Operating Temperature........................................... 15 °C to 40 °C
Storage Temperature.............................................. -20 °C to +65 °C
Relative Humidity .................................................... 90 % max