Remote Operation

Programming with Analyzer Commands

3

3-45

DEMO ROOM

Page: 1

Date: 06/05/93

Time: 12:32 AM

Proc. Name

SENT*ADULT

Description

GENERAL CONDITION OF EQUIPMENT

LINE CORD/STRAIN RELIEF

TRPN (POWER ON)

INDICATORS/DISPLAYS

CONTROLS/SWITCHES

TUBES/HOSES

FITTINGS/CONNECTORS

IDENT

LEAKTEST

LEAK TEST: APPLY 200 MMHG (<10/MIN)

MANOMETER

STATIC PRESSURE CHECK: APPLY 200 MMHG

STATIC PRESSURE CHECK: APPLY 100 MMHG

POP OFF/RELIEF VALVE: APPLY >250 MMHG

MAKEARM:HR=100;BP=ADAMS-120/80

ADULT 120/80 AT 100 BPM

MAKEARM:HR=80;BP=ADAMS-150/100

ADULT 150/100 AT 80 BPM

MAKEARM:HR=60;BP=ADAMS-200/150

ADULT 200/150 AT 60 BPM

GOTOLOCAL

GTOL

5

GEN/PHYS

GEN/MECH

REMOTE

GEN/ELEC

GEN/ELEC

GEN/PLMB

GEN/PHYS

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

AUTO SEQ

Pass/Fail

Pass/Fail

medTester remote command

Pass/Fail

Pass/Fail

Pass/Fail

Pass/Fail

medTester remote command

medTester remote command

Pass/Fail

medTester remote command

Pass/Fail

Pass/Fail

Pass/Fail

medTester remote command

Pass/Fail

medTester remote command

Pass/Fail

medTester remote command

Pass/Fail

medTester remote command

medTester remote command

medTester autosequence

Type

medTester Control

0.00

PM Labor Time

Relational Checklist Report

Listed In Order Of Relational Procedure

fcv067.eps

Figure 3-9. Sentinel Adult BP Checklist