Makearmneo – Fluke Biomedical Cufflink User Manual

Page 97

Advertising
background image

Remote Operation

Command Descriptions

3

3-23

Example

Simulate a blood pressure of 120/80 with a heart rate of 30 BPM, a gain of 50%, and a shift of
12 mmHg.

Command

makearm

bp=adams-120/80,hr=30,

envgain=50,envshift=12

Results

The Analyzer draws the makearm graph and begins
arm simulation when the NIBP monitor starts inflating
the cuff. Test results are sent to the controller when the
cuff deflates.

C

u

ffPres

C

u

ffPeak

DeflRate
DeflTime
InflRate
InflTime
TotlTime
120/

8

0

0
142
3.1
45.1
43.2
3.1
4

8

.1

(90) A

200

160

120

40

0

20

40

60

30 BPM

G 50
S 12

8

0

fcv049.eps

Returned Data

142, 3.1, 45.1, 43.2, 3.1, 48.1; varies, depending on
parameter values

MAKEARMNEO

Use this command to simulate blood pressures from the ADAMS Neonate Family target

values. Heart rate, BP, and pressure envelope gain and shift can be modified. Parameters

for this command are described below.

Parameter Definition

Range Default

Description

HR Heart

Rate

(Beats/Minute)

30 BPM to 240
BPM in 1 BPM
Steps

80 BPM

1 BPM resolution

BP Blood

Pressure

(mmHg)

ADAMSNEO-
60/30

ADAMSNEO-
80/50

ADAMSNEO-
100/65

ADAMSNEO-
120/80

ADAMSNEO-
150/100

ADAMSNEO-
200/150

ADAMSNEO-
255/195

ADAMSNEO-
120/80