Fluke Biomedical Cufflink User Manual
Page 96
Advertising

Cufflink
Operators Manual
3-22
Peak Press
u
re
(mmHg)
(mmHg/min)
(min
u
tes)
(min
u
tes)
(mmHg/min)
(min
u
tes)
Deflate Time
Deflate Rate
Inflate Time
Inflate Rate
Total Time
79,
4.0,
2.0,
7
8
.9,
0.9,
3.3
fcv047.eps
Example
Simulate a blood pressure of 100/65 with a heart rate of 60 BPM.
Command
makearm
bp=adams-100/65,hr=60
Results
The Analyzer draws the makearm graph and begins
arm simulation when the NIBP monitor starts inflating
the cuff. Test results are sent to the controller when the
cuff deflates.
C
u
ffPres
C
u
ffPeak
DeflRate
DeflTime
InflRate
InflTime
TotlTime
120/
8
0
0
197
4.9
39.1
39.4
4.
8
43.
8
(75) A
200
160
120
40
0
20
40
60
60 BPM
8
0
fcv048.eps
Returned Data
197, 4.9, 39.1, 39.4, 4.8, 43.8; varies, depending on
parameter values
Advertising