Remote commands, General safety considerations, Symbols – Fluke Biomedical Cufflink User Manual

Page 18

Advertising
background image

Cufflink

Operators Manual

1-6

Remote Commands

Table 1-1 lists new CuffLink RS232 commands to support new functions.

Table 1-1. New RS232 Remote Commands

Command Function

DEFLATE

Releases pressure inside CuffLink

INFLATE

Pumps CuffLink to 200 mmHg or specified pressure

KEYTEST

Tests CuffLink's keyboard

MKARR_AF

Simulates atrial fibrillation

MKARR_MB

Simulates a missed beat

MKARR_PAC

Simulates premature atrial contraction

MKARR_PVC

Simulates premature ventricular contraction

MKARR_AS

As Simulates aberrant sinus conduction

POPOFF

Tests monitor's overpressure valve

PUMPPCB

Determines if pump PCB is installed in CuffLink

General Safety Considerations

Read the Users Manual before operating the Analyzer.

Symbols

Table 1-2 describes the symbols associated with the Analyzer.

Table 1-2. Symbols

Symbol Description

X

Hazardous voltage

W

Important information; refer to manual.

Œ

Conforms to UL Std 3101-1; certified to CAN/USA Std C22.2 No.
1010.1

P

Conforms to European Union directives

~

Do not dispose of this product as unsorted municipal waste. Go
to Fluke’s website for recycling information.

Hg

Contains mercury. Dispose properly.