Adult arrhythmic bp checklists and test results, Remote operation – Fluke Biomedical Cufflink User Manual

Page 125

Advertising
background image

Remote Operation

Programming with Analyzer Commands

3

3-51

Adult Arrhythmic BP Checklists and Test Results

Figure 3-15 shows a medBase Adult BP Arrhythmic checklist that creates results like

those shown in Figure 3-16. Figure 3-17 shows a Sentinel Adult Arrhythmic BP checklist

that creates results like those shown in Figure 3-18.

DATE: 06/05/1993

CHECKLIST NUMBER: 17 (start)

Ln

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

17

18

19

20

21

22

23

80

1

2

37

6

7

13

15

38

39

40

41

42

43

73

44

45

46

47

48

49

50

51

----->

25

\P GENERAL CONDITION OF EQUIPMENT

\P LINE CORD/STRAIN RELIEF

\M TRPN (POWER ON)

\P INDICATORS/DISPLAYS

\P CONTROLS/SWITCHES

\P TUBES/HOSES

\P FITTINGS/CONNECTORS

\M IDENT

\M LEAKTEST

\P LEAK TEST: APPLY 200 NMHG (<10/MIN)

\M MANOMETER

\P STATIC PRESSURE CHECK: APPLY 200 MMHG

\P STATIC PRESSURE CHECK: APPLY 100 MMHG

\P POP OFF/RELIEF VALVE: APPLY >250 MMHG

\M MKARR_PAC PRE. ATRIAL CONTRACTION

\P ARRHYTHMIA PAC 120/80 @ 80 BPM

\M MKARR_PVC PRE. VENTRICULAR CONTRACTION

\P ARRHYTHMIA PVC 120/80 @ 80 BPM

\M MKARR_AF ATRIAL FIBRILLATION

\P ARRHYTHMIA ATRIAL FIB 120/80 @ 80 BPM

\M GOTOLOCAL

\M GTOL

through line 79 - unused -

-A4

GEN/PHYS

GEN/MECH

REMOTE

GEN/ELEC

GEN/ELEC

GEN/PLMB

GEN/PHYS

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

AUTO SEQ

Item

Check Item

Type

CHECKLIST DESCRIPTION: NIBP*ARR

CHECKLIST CONTENTS

PAGE: 2

--- end of checklist #17 --

fcv074.eps

Figure 3-15. medBase Adult Arrhythmic BP Checklist

Advertising