Non-Invasive Blood Pressure (NIBP) Monitoring Tutorial
A
A-3
Oscillometric Response
(Press
u
re P
u
lses)
A
u
sc
u
ltatory Response
(Korotkoff So
u
nds)
C
u
ff Press
u
re
MAP
MAP
Distole
Distole
Systole
Systole
Time
Time
Time
fcv006.eps
Figure A-1. Auscultatory vs. Oscillometric Techniques
Also in Figure A-1, the oscillometric pulse amplitudes are depicted as constantly
changing. There is no point where the pulses abruptly change in amplitude, but there is a
point where the pulse amplitude reaches a peak. It is generally agreed that the cuff
pressure at peak pulse amplitude is the MAP, a value not easily identified using the
auscultatory method.