Cufflink

Operators Manual

3-50

MedTester REC # 18

CHECKLIST: NIBP*NEO

OP CODE: TEO

DEVICE INFORMATION

PHYSICAL INSPECTION

COMMENTS:

NEXT TEST DUE DATE:

USER TIME:

ELAPSED TEST TIME: 263 SECS

GENERAL CONDITION OF EQUIPMENT........PASS

LINE CORD/STRAIN RELIEF...............PASS

TRPN (POWER ON).......................REMOTE CMD, NO DATA

INDICATORS/DISPLAYS...................PASS

CONTROLS/SWITCHES.....................PASS

TUBES/HOSES...........................PASS

FITTINGS/CONNECTORS...................PASS

IDENT.................................CUFFLINK,3.00,AUSC,PUMP

LEAKTEST..............................*

LEAK TEST: APPLY 200 MMHG (<10/MIN)...PASS

MANOMETER.............................*

STATIC PRESSURE CHECK: APPLY 200 MMHG.PASS

STATIC PRESSURE CHECK: APPLY 100 MMHG.PASS

POP OFF/RELIEF VALVE: APPLY >250 MMHG.PASS

MAKEARMNEO:HR=60;BP=ADAMSNEO-150/100.....*

ADAMS NEONATE 150/100 AT 60 BPM.......PASS

MAKEARMNEO:HR=120;BP=ADAMSNEO-100/65..*

ADAMS NEONATE 100/65 AT 120 BPM.......PASS

MAKEARMNEO:HR=200;BP=ADAMSNEO-60/30...*

ADAMS NEONATE 60/30 AT 200 BPM........PASS

GOTOLOCAL.............................*

GTOL..................................REMOTE CMD, NO DATA

TYPE: NIBP*NEO

MODEL:

MANF:

SN:

LOC: DNI

CN: NIBP*NEO

DATE: 6/05/93

TIME: 15:44:19

fcv071.eps

Figure 3-14. Sentinel Neonate BP Test Results