Fluke Biomedical Cufflink User Manual

Page 25

Advertising
background image

Introduction and Specifications

Specifications

1

1-13

Display

Alphanumeric and Graphic LCD Display

Alphanumeric Mode............................................ 8 lines by 40 characters
Graphics Mode ................................................... 64 vertical by 240 horizontal dot matrix

Illumination.............................................................. Backlight with Viewing Angle Adjustment
Displayed Graphics................................................. Dynamic real-time NIBP cuff-pressure waveform, programmed
peripheral pulse and envelope waveforms

Control Keys

Function Keys ........................................................ F1 to F5

Enter
Escape

Cursor .................................................................... Up

Down
Left
Right

Parameter Selections

Menus ..................................................................... Pull down with on-screen help
Function keys.......................................................... Software defined, F1 to F5

Storage

Environment............................................................ Store in a dry area, temperature range of 32°F to 122°F.
Inspection or maintenance during storage.............. None required

Digital Interfaces

RS232/Serial........................................................... Baud Rates: 300, 600, 1200, 2400, 4800, 9600

Stop Bits: 1, 2
Parity: Odd, Even, Off
Handshake: Xon/Xoff, RTS/CTS, None

Parallel Port (Printer) .............................................. Centronics compatible
Pulse Sync .............................................................. 0 to 5 V dc (TTL)

Target Blood Pressure Selections

Select BP

Target Value (MAP)

(mmHg)

HR

(BPM)

Waveform Type

Adult 60/30

(40)

80/50 (62)
100/65 (75)
120/80 (90)
150/100 (115)
200/150 (165)
255/195 (215)

30
40
60

120
160
200
240

Normal Sinus

Neonate 60/30

(40)

80/50 (62)
100/65 (75)
120/80 (90)
150/100 (115)

30
40
60

120
160
200
240

Normal Sinus

Arrhythmias 120/80

(90)

80

Premature Atrial Contraction
Premature Ventricular Contraction
Atrial Fibrillation
Missed Beat
Aberrant Sinus Conduction

Advertising