Fluke Biomedical Cufflink User Manual

Page 73

Advertising
background image

Operation

Printing Documents

2

2-45

NEO-2

Page 1

Date:09/25/93

Time: 08:23

Manufacturer:____________

ABC Co

YYY

00104

N200

PICU

5.25

BED 2

3.0

Control#: ________

Software Rev:____________

Leak Start: 337.00 mmHg

Pop Off: 263.25 mmHg

Blood

Pressure

100/63 (77)

120
120

195

Gain=100

6.0

31.1

49.2

3.8

34.9

Shift=0

100/65 (75)A

Heart

Rate

(BPM)

Peak

Press

(mmHg)

Deflate

Rate

(mmHg/s)

Deflate

Time

(sec)

Inflate

Rate

(mmHg/s)

Inflate

Time

(sec)

Total

Time

(sec)

Leak Rate: mmHg/m

***** Leak Test *****

***** Over Pressure Test *****

***** Make Arm Results *****

Model:____________

Dept :___________

Ser:____________

Loc:____________

Notes:________________________________

Cufflink Auto Sequence

98/63 (76)

122
120

100/65 (75)A

099/63 (79)

122
120

100/65 (75)A

98/ 64 (76)

205
200

141

Gain=100

10.1

13.2

51.2

2.6

15.8

Shift=0

NOTE:Bold type represents data hand-entered by user

100/ 65 (75)A

100/ 64 (78)

205
200

140

Gain=100

8.1

10.9

45.5

2.9

15.1

Shift=0

100/ 65 (75)A

99/ 65 (75)

205
200

149

Gain=100

10.6

13.2

23.9

5.9

19.1

Shift=0

100/ 65 (75)A

59/28 (43)

97

Gain=100

6.7

13.1

38.7

2.3

15.4

Shift=0

60/30 (40)A

59/29 (39)

94

Gain=100

6.1

14.1

53.8

1.6

15.7

Shift=0

60/30 (40)A

59/29 (39)

122

Gain=100

6.5

17.5

42.2

2.7

20.2

Shift=0

60/30 (40)A

79/49 (62)

122

Gain=100

5.3

16.7

54.3

2.1

19.5

Shift=0

80/50 (62)A

80/49 (61)

121

Gain=100

6.6

17.4

56.5

2.0

19.4

Shift=0

80/50 (62)A

79/48 (59)

120
120

135

Gain=100

5.8

22.1

45.4

2.8

24.9

Shift=0

80/50 (62)A

122
120

120
120

163
160

163
160

160
160

138

Gain=100

5.9

16.7

65.0

2.0

19.9

Shift=0

137

Gain=100

5.2

18.7

61.4

2.1

22.0

Shift=0

fcv090.eps

Figure 2-16. Sample Printout of Neonate Auto Sequence Test Results