Fluke Biomedical Cufflink User Manual

Page 127

Advertising
background image

Remote Operation

Programming with Analyzer Commands

3

3-53

DEMO ROOM

Page: 1

Date: 06/05/93

Time: 12:34 AM

Proc. Name

SENT*ARR

Description

GENERAL CONDITION OF EQUIPMENT

LINE CORD/STRAIN RELIEF

TRPN (POWER ON)

INDICATORS/DISPLAYS

CONTROLS/SWITCHES

TUBES/HOSES

FITTINGS/CONNECTORS

IDENT

LEAKTEST

LEAKTEST: APPLY 200 MMHG (<10/MIN)

MANOMETER

STATIC PRESSURE CHECK: APPLY 200 MMHG

STATIC PRESSURE CHECK: APPLY 100 MMHG

POP OFF/RELIEF VALVE: APPLY >250 MMHG

MKARR_PAC PRE. ATRIAL CONTRACTION

AARRHYTHMIA PAC 120/80 AT 80 BPM

MKARR_PVC PRE VENTRICULAR CONTRACTION

AARRHYTHMIA PVC 120/80 AT 80 BPM

MKARR_AF ATRIAL FIBRILLATION

AARRHYTHMIA ATRIAL FIB 120/80 AT 80 BPM

GOTOLOCAL

GTOL

5

GEN/PHYS

GEN/MECH

REMOTE

GEN/ELEC

GEN/ELEC

GEN/PLMB

GEN/PHYS

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

CL

AUTO SEQ

Pass/Fail

Pass/Fail

medTester remote command

Pass/Fail

Pass/Fail

Pass/Fail

Pass/Fail

medTester remote command

medTester remote command

Pass/Fail

medTester remote command

Pass/Fail

Pass/Fail

Pass/Fail

medTester remote command

Pass/Fail

medTester remote command

Pass/Fail

medTester remote command

Pass/Fail

medTester remote command

medTester remote command

medTester autosequence

Type

medTester Control

0.00

PM Labor Time

Relational Checklist Report

Listed In Order Of Relational Procedure

fcv079.eps

Figure 3-17. Sentinel Adult Arrhythmic BP Checklist

Advertising