Manometer, Mkarm10, Mkarm20 – Fluke Biomedical Cufflink User Manual

Page 101

Advertising
background image

Remote Operation

Command Descriptions

3

3-27

MANOMETER

Use this command to put the Analyzer into the manometer mode. The keys on the

Analyzer front panel become active during this test so a user can zero pressure, do a reset,

or print test results. Press F1 to start measurements. See also: Pop Off Test. There are no

parameters for this command.

Example

Put CuffLink into the manometer mode.

Command

manometer

Results

The Analyzer enters the manometer test mode and
shows the Manometer display. Press Fl on the
Analyzer keyboard to begin.

fcv056.eps

Returned

Data

*

MKARM10

Use this command to set blood pressure simulation to Adams Adult 60/30, with a heart

rate of 40 BPM. There are no parameters for this command.

Example

Simulate a specific adult BP of 60/30 at 40 BPM.

Command Mkarm10

Results

The Analyzer runs the Adams Adult 60/30, heart rate is set to 40
BPM.

Returned

Data

*

MKARM20

Use this command to set blood pressure simulation to Adams Adult 80/50, with a heart

rate of 60 BPM. There are no parameters for this command.

Example

Simulate a specific adult BP of 80/50 at 60 BPM.

Command Mkarm20

Results

The Analyzer runs the Adams Adult 80/50, heart rate is set to 60
BPM.

Returned

Data

*

Advertising