Pulse – Fluke Biomedical Cufflink User Manual

Page 108

Advertising
background image

Cufflink

Operators Manual

3-34

Example

Output the default pulse.

Command

pscale

Results

Pulse is output at PULSE output jack on the Analyzer.

F1

F2

F3

F4

F5

*** Scale Pulse Amplitude ***

Motion : Running Enter toggles
Amplitude : 1000 motion start/
Pressure

: 0.00 stop

Rate (BPM) 80

Rate

Shape 250

500

1000

fcv059.eps

Returned

Data

*

Example

Output a 28 millisecond pulse at 40 BPM with an amplitude of 500 RVDUs.

Command

pscale

hr=40,pulseid=5,pulseamp=500

Results

Pulse is output at PULSE output jack on the Analyzer.

F1

F2

F3

F4

F5

*** Scale Pulse Amplitude ***

Motion : Running Enter toggles
Amplitude : 500 motion start/
Pressure

: 0.00 stop

Rate (BPM) 40

Rate

Shape 250

500

1000

fcv060.eps

Returned

Data

*

PULSE

Use this command to output a single, specific pressure pulse. If parameters are not

specified, the Analyzer uses a current value for pulse width and calculates pulse

amplitude. The Analyzer calculates pulse amplitude by reading the cuff pressure and

applying it to the active pressure envelope.
This command is useful for synchronizing Analyzer pulse output to another device and

for custom pressure envelopes with random pulse selection. Parameters for this command

are described below.