Mkarmneo40, Mkarmneo50, Mkarr_af – Fluke Biomedical Cufflink User Manual

Page 104: Mkarr_mb

Advertising
background image

Cufflink

Operators Manual

3-30

MKARMNEO40

Use this command to set blood pressure simulation to Adams Neonate 120/80, with a

heart rate of 120 BPM. There are no parameters for this command.

Example

Simulate a specific neonate BP of 120/80 at 120 BPM.

Command mkarmneo40

Results

The Analyzer runs the Adams Neonate 120/80, heart rate is set to 120
BPM.

Returned

Data

*

MKARMNEO50

Use this command to set blood pressure simulation to Adams Neonate 150/100, with a

heart rate of 120 BPM. There are no parameters for this command.

Example

Simulate a specific neonate BP of 150/100 at 120 BPM.

Command mkarmneo50

Results

The Analyzer runs the Adams Neonate 150/100, heart rate is set to
120 BPM.

Returned

Data

*

MKARR_AF

Use this command to set blood pressure simulation to the Atrial Fibrillation Arrhythmia

at the fixed BP and heart rate of 120/80 and 80 BPM.

There are no parameters for this

command.

Example

Simulate an atrial fibrillation.

Command

mlcarr_af

Results

The Analyzer runs atrial fibrillation at 120/80 with the
heart rate set to 80 BPM.

Returned

Data

*

MKARR_MB

Use this command to set blood pressure simulation to the Missed Beat Arrhythmia at the

fixed BP and heart rate of 120/80 and 80 BPM. There are no parameters for this

command.

Example

Simulate a missed beat.

Command mitarr_mb

Results

The Analyzer runs missed beat at 120/80 with the heart rate set to 80
BPM.

Returned

Data

*