Preliminary procedures, Assembling equipment – Fluke Biomedical Cufflink User Manual

Page 34

Advertising
background image

Cufflink

Operators Manual

2-6

Preliminary Procedures

The analyzer simulates a human arm and produces target value blood pressures for the

purpose of testing the accuracy of blood pressure readings on an NIBP monitor. The test

is initiated from the

Select BP

Main Menu

option that provides the parameters listed in

Table 2-1.

Table 2-1. Blood Pressure Parameters Provided by the Select BP Option

Adult Neonate

Adult

Arrhythmias

Blood

Pressure

(mmHg)

Mean Arterial

Pressure (mmHg)

Blood Pressure

(mmHg)

Mean Arterial

Pressure (mmHg)

60/30
80/50

400/65
120/80

150/100
200/150
255/195

40
62
75
90

115
165
215

60/30
80/50

400/65
120/80

150/100

40
62
75
90

115

Heart Rate Selections

Premature Atrial
Contractions

Premature
Ventricular
Contraction

Atrial fibrillation

Missed Beat

Aberrant Sinus
Conductions

Blood Pressure

and Heart Rate

30
40
60
80

120
160
200
240

Blood Pressure
fixed at 120/80 (90)

Heart Rate fixed at
80 BPM

Assembling Equipment

The following equipment is needed to test NIBP monitor blood pressure readings. Figure

2-2 shows the proper way to combine Adult cuff mandrels to simulate various sizes of

arms.

NIBP monitor

NIBP Analyzer

Mandrel (supplied with the Analyzer)

BP cuff

Hoses to attach the cuff to the monitor

Cuff adapter for the monitor DUT (device under test) (supplied with the Analyzer)