Cufflink

Operators Manual

3-46

MedTester REC # 16

CHECKLIST: ADULT*NIBP

OP CODE:TEO

DEVICE INFORMATION

PHYSICAL INSPECTION

COMMENTS:

NEXT TEST DUE DATE:

USER TIME:

ELAPSED TEST TIME: 209 SECS

GENERAL CONDITION OF EQUIPMENT........PASS

LINE CORD/STRAIN RELIEF...............PASS

TRPN (POWER ON).......................REMOTE CMD, NO DATA

INDICATORS/DISPLAYS...................PASS

CONTROLS/SWITCHES.....................PASS

TUBES/HOSES...........................PASS

FITTINGS/CONNECTORS...................PASS

IDENT.................................CUFFLINK,3.00,AUSC,PUMP

LEAKTEST..............................*

LEAK TEST: APPLY 200 MMHG (<10/MIN)...PASS

MANOMETER.............................*

STATIC PRESSURE CHECK: APPLY 200 MMHG.PASS

STATIC PRESSURE CHECK: APPLY 100 MMHG.PASS

POP OFF/RELIEF VALVE: APPLY >250 MMHG.PASS

MAKEARM:HR=100;BP=ADAMS-120/80........*

ADULT 120/80 AT 100 BPM...............PASS

MAKEARM:HR=80;BP=ADAMS-150/100........*

ADULT 150/100 AT 80 BPM...............PASS

MAKEARM:HR=60;BP=ADAMS-200/150........*

ADULT 200/150 AT 60 BPM...............PASS

GOTOLOCAL.............................*

GTOL..................................REMOTE CMD, NO DATA

TYPE: ADULT*NIBP

MODEL:

MANF:

SN:

LOC: DNI

CN: ADULT*NIBP

DATE: 6/05/93

TIME: 15:39:31

fcv064.eps

Figure 3-10. Sentinel Adult BP Test Results